Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Coriolis flowmeter - Proline Promass 80H

Promass 80H

80H
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Coriolis

  • Product headline

    The chemically resistant single‐tube flowmeter with a compact or remote transmitter. Highly accurate measurement of liquids and gases in applications requiring highest corrosion resistance.

  • Sensor features

    Maximum safety for chemically aggressive fluids – corrosionresistant wetted parts. Fewer process measuring points – multivariable measurement (flow, density, temperature). Spacesaving installation – no in/outlet run needs. Measuring tube made of Tantalum, Zirconium. Nominal diameter: DN 8 to 50 (³⁄₈ to 2").

  • Transmitter features

    Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.

  • Nominal diameter range

    DN 8 to 50 (³⁄₈ to 2")

  • Wetted materials

    Measuring tube: Tantalum 2.5W; 702 (UNS R60702)
    Connection: Tantalum; 702 (UNS R60702)

  • Measured variables

    Mass flow, density, temperature, volume flow, corrected volume flow, reference density

  • Max. measurement error

    Mass flow (liquid): ±0.15 %
    Volume flow (liquid): ±0.15 %
    Mass flow (gas, Tantalum only): ±0.5 %
    Density (liquid): ±0.0005 g/cm³

  • Measuring range

    0 to 70 000 kg/h (0 to 2570 lb/min)

  • Max. process pressure

    PN 40, Class 300, 20K

  • Medium temperature range

    Tantalum: –50 to +150 °C (–58 to +302 °F)
    Ziconium: –50 to +200 °C (–58 to +392 °F)

  • Ambient temperature range

    Standard: –20 to +60 °C (–4 to +140 °F)
    Option: –40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

    1.4301 (304), corrosion resistant

  • Transmitter housing material

    Powder‐coated die‐cast aluminium
    1.4301 (304), sheet
    CF3M (316L), cast

  • Degree of protection

    IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure

  • Display/Operation

    2‐line backlit display with push buttons
    Configuration via local display and operating tools possible

  • Outputs

    3 outputs:
    0‐20 mA (active)/4‐20 mA (active/passive)
    Pulse/frequency/switch output (passive)

  • Inputs

    Status input

  • Digital communication

    HART
    PROFIBUS PA

  • Power supply

    DC 16 to 62 V
    AC 85 to 260 V (45 to 65 Hz)
    AC 20 to 55 V (45 to 65 Hz)

  • Hazardous area approvals

    ATEX, IECEx, FM, CSA, NEPSI

  • Other approvals and certificates

    3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
    PED, CRN

  • Product safety

    CE, C-tick, EAC marking

  • Functional safety

    Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

    Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

  • Pressure approvals and certificates

    PED, CRN

  • Material certificates

    3.1 material

  • Measuring principle

    Coriolis

  • Product headline

    The chemically resistant single‐tube flowmeter with a compact or remote transmitter. Highly accurate measurement of liquids and gases in applications requiring highest corrosion resistance.

  • Sensor features

    Maximum safety for chemically aggressive fluids – corrosionresistant wetted parts. Fewer process measuring points – multivariable measurement (flow, density, temperature). Spacesaving installation – no in/outlet run needs. Measuring tube made of Tantalum, Zirconium. Nominal diameter: DN 8 to 50 (³⁄₈ to 2").

  • Transmitter features

    Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.

  • Nominal diameter range

    DN 8 to 50 (³⁄₈ to 2")

  • Wetted materials

    Measuring tube: Tantalum 2.5W; 702 (UNS R60702)
    Connection: Tantalum; 702 (UNS R60702)

  • Measured variables

    Mass flow, density, temperature, volume flow, corrected volume flow, reference density

  • Max. measurement error

    Mass flow (liquid): ±0.15 %
    Volume flow (liquid): ±0.15 %
    Mass flow (gas, Tantalum only): ±0.5 %
    Density (liquid): ±0.0005 g/cm³

  • Measuring range

    0 to 70 000 kg/h (0 to 2570 lb/min)

  • Max. process pressure

    PN 40, Class 300, 20K

  • Medium temperature range

    Tantalum: –50 to +150 °C (–58 to +302 °F)
    Ziconium: –50 to +200 °C (–58 to +392 °F)

  • Ambient temperature range

    Standard: –20 to +60 °C (–4 to +140 °F)
    Option: –40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

    1.4301 (304), corrosion resistant

  • Transmitter housing material

    Powder‐coated die‐cast aluminium
    1.4301 (304), sheet
    CF3M (316L), cast

  • Degree of protection

    IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure

  • Display/Operation

    2‐line backlit display with push buttons
    Configuration via local display and operating tools possible

  • Outputs

    3 outputs:
    0‐20 mA (active)/4‐20 mA (active/passive)
    Pulse/frequency/switch output (passive)

  • Inputs

    Status input

  • Digital communication

    HART
    PROFIBUS PA

  • Power supply

    DC 16 to 62 V
    AC 85 to 260 V (45 to 65 Hz)
    AC 20 to 55 V (45 to 65 Hz)

  • Hazardous area approvals

    ATEX, IECEx, FM, CSA, NEPSI

  • Other approvals and certificates

    3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
    PED, CRN

  • Product safety

    CE, C-tick, EAC marking

  • Functional safety

    Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

    Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

  • Pressure approvals and certificates

    PED, CRN

  • Material certificates

    3.1 material

  • Measuring principle

    Coriolis

  • Characteristic / Application

    Balanced single-tube system, "Fit-and-Forget" installation
    Design:
    Easy to clean, hygienic, careful handling of the medium
    - chemically resistant material

  • Ambient temperature

    -20...+65°C
    (-4...+140°F)

  • Process temperature

    -50...+200°C
    (-58...+392°F)

  • Process pressure

    PN 40
    Cl 150...300
    JIS 10...20K

  • Wetted parts

    Zirconium 702/R60702

  • Output

    4...20 mA
    Pulse/Frequency
    Status

  • Certificates / Approvals

    ATEX
    FM
    CSA

  • Measuring principle

    Coriolis

  • Product headline

    The chemically resistant single‐tube flowmeter with a compact or remote transmitter. Highly accurate measurement of liquids and gases in applications requiring highest corrosion resistance.

  • Sensor features

    Maximum safety for chemically aggressive fluids – corrosionresistant wetted parts. Fewer process measuring points – multivariable measurement (flow, density, temperature). Spacesaving installation – no in/outlet run needs. Measuring tube made of Tantalum, Zirconium. Nominal diameter: DN 8 to 50 (³⁄₈ to 2").

  • Transmitter features

    Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.

  • Nominal diameter range

    DN 8 to 50 (³⁄₈ to 2")

  • Wetted materials

    Measuring tube: Tantalum 2.5W; 702 (UNS R60702)
    Connection: Tantalum; 702 (UNS R60702)

  • Measured variables

    Mass flow, density, temperature, volume flow, corrected volume flow, reference density

  • Max. measurement error

    Mass flow (liquid): ±0.15 %
    Volume flow (liquid): ±0.15 %
    Mass flow (gas, Tantalum only): ±0.5 %
    Density (liquid): ±0.0005 g/cm³

  • Measuring range

    0 to 70 000 kg/h (0 to 2570 lb/min)

  • Max. process pressure

    PN 40, Class 300, 20K

  • Medium temperature range

    Tantalum: –50 to +150 °C (–58 to +302 °F)
    Ziconium: –50 to +200 °C (–58 to +392 °F)

  • Ambient temperature range

    Standard: –20 to +60 °C (–4 to +140 °F)
    Option: –40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

    1.4301 (304), corrosion resistant

  • Transmitter housing material

    Powder‐coated die‐cast aluminium
    1.4301 (304), sheet
    CF3M (316L), cast

  • Degree of protection

    IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure

  • Display/Operation

    2‐line backlit display with push buttons
    Configuration via local display and operating tools possible

  • Outputs

    3 outputs:
    0‐20 mA (active)/4‐20 mA (active/passive)
    Pulse/frequency/switch output (passive)

  • Inputs

    Status input

  • Digital communication

    HART
    PROFIBUS PA

  • Power supply

    DC 16 to 62 V
    AC 85 to 260 V (45 to 65 Hz)
    AC 20 to 55 V (45 to 65 Hz)

  • Hazardous area approvals

    ATEX, IECEx, FM, CSA, NEPSI

  • Product safety

    CE, C-tick, EAC marking

  • Functional safety

    Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

    Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

  • Pressure approvals and certificates

    PED, CRN

  • Material certificates

    3.1 material